ASTM F136 Titanium: What It Means and How to Verify It
What F136 fixes, how to read a 3.1 report, and a worked example built on real certificates.
The Tremun Team
Two bars of titanium can look identical and polish to the same colour. One is ASTM F136. The other is industrial Grade 5. Only a document tells them apart. This guide shows what the standard fixes, how to read that document, and when to refuse it.
We are a titanium supplier, so the worked example uses our own published documents. A mill report describes the bar. A lab test on finished pieces checks the jewellery. We publish both, so you can apply every check in this guide to us first.
For metal comparisons, read Titanium is better than steel y Why titanium is the best material for piercing jewellery.
Definition
ASTM F136 is the standard for Ti-6Al-4V ELI, a titanium alloy made for surgical implants. ELI means the alloy keeps oxygen and other small atoms very low. The standard sets chemistry limits, such as oxygen ≤0.13%, and minimum strength. It describes the metal bar, not the finished jewellery.
In short
Five checks before you accept a titanium supplier
- The standard is named. ASTM F136 or ISO 5832-3. Not only “titanium”, “G23” or “implant grade”.
- The certificate is type 3.1. An EN 10204 3.1 report, signed and dated.
- The heat number matches. The same number appears on every page.
- The numbers sit inside the limits. For F136: oxygen ≤0.13% (ISO 5832-3 allows up to 0.20%), iron ≤0.25%, tensile strength ≥860 MPa.
- The finished piece passes your loupe. Mirror surfaces, clean threads, no sharp edges.
Is F136 implant grade titanium good for piercings?
Yes, with conditions. ASTM F136 is one of the titanium standards in the APP minimum standard for initial jewellery. But the letters “F136” on a website prove nothing alone. The claim is only as good as the certificate behind it.
Design, polish and threads come later, in the workshop. A good certificate cannot rescue a poor finish. A good finish cannot replace a missing certificate.
What is ASTM F136? What the standard specifies
ASTM F136 is a material specification for Ti-6Al-4V ELI bar and wire for surgical implants. Its alloy code is UNS R56401, and the trade often calls it “Grade 23”. The standard fixes the chemistry, the minimum strength and the annealed structure of the metal.
Read the oxygen line first
Oxygen is the main difference between F136 and ordinary Ti-6Al-4V, sold as Grade 5. F136 allows a maximum of 0.13% oxygen. Grade 5 allows up to 0.20%, and so does ISO 5832-3. Only the certificate shows which limit your metal meets.
| Requirement | ASTM F136 limit |
|---|---|
| Aluminium (Al) | 5.50–6.50% |
| Vanadium (V) | 3.50–4.50% |
| Iron (Fe) | ≤0.25% |
| Oxygen (O) | ≤0.13% |
| Carbon (C) | ≤0.08% |
| Nitrogen (N) | ≤0.05% |
| Hydrogen (H) | ≤0.012% |
| Tensile strength | ≥860 MPa |
| Yield strength | ≥795 MPa |
| Elongation | ≥10% |
Chemistry in weight per cent. Hydrogen may reach 0.015% in very thin material (0.813 mm or less). Mechanical minimums for annealed bar and wire in jewellery sizes, such as 4 mm bar.
Editions: what the year means
The year after the number is the edition, such as F136-13 or F136-26. F136-26 has been the active edition since February 2026. A mill report cites the edition in force when the bar was made, so older editions remain common. Check that an edition is named, and note it in your records.
F136 vs G23, Grade 5 and ISO 5832-3
“Implant grade” is not a standard at all. It only counts when a named specification follows it.
| Name on the document | Alloy | Oxygen max. | APP initial jewellery |
|---|---|---|---|
| ASTM F136 | Ti-6Al-4V ELI | 0.13% | Yes |
| ISO 5832-3 | Ti-6Al-4V | 0.20% | Yes |
| ASTM F1295 | Ti-6Al-7Nb | - | Yes |
| ASTM F67 | Unalloyed titanium | Depends on grade (1–4) | Yes |
| ASTM B348 Grade 23 | Ti-6Al-4V ELI | 0.13% | Not by name |
| ASTM B348 Grade 5 | Ti-6Al-4V | 0.20% | No |
| “Implant grade” | No standard behind it | - | Only with a named standard |
Decision rules from this table:
- F136 named: accept, then check the values against the F136 limits above.
- ISO 5832-3 named: accept, then check the values against ISO 5832-3 limits. ISO 5832-3 is not ELI. It allows oxygen up to 0.20% and iron up to 0.30%. F136 limits are tighter for oxygen, iron and yield strength.
- Only “Grade 23” or “G23”: not enough. ASTM B348 Grade 23 is the same alloy family under an industrial bar specification, with slightly different limits. It has no implant requirements and a different alloy code, UNS R56407. Request the implant standard.
- “Grade 5” or “Ti-6Al-4V” with no implant standard: reject for initial piercings. ASTM B348 Grade 5 is an industrial bar specification, and APP does not list it.
One report can name several standards at once. A single bar can meet F136, ISO 5832-3 and B348 Grade 23 together. That is normal, and you will see it in our own reports below. What matters is that an implant standard is on the list.
Why EN 10204 3.1, not 2.2
The chain. Each melt of titanium gets a heat number. That melt becomes bar lots, and a bar lot becomes jewellery batches. The heat number links the paper to the metal. Without it, a certificate is only a general statement.
The test report for that melt is an inspection document, often called a mill certificate. The European standard EN 10204 defines four types.
No
Type
2.1
A declaration of compliance. No test results.
No
Type
2.2
General test results for the product type, not your heat number.
Yes, the normal level
Type
3.1
Test results for your heat number or lot. Signed by the mill’s authorised inspector, who works outside production.
Yes, rare for jewellery
Type
3.2
A 3.1, also signed by the buyer’s inspector or an independent inspector.
In short, request a 3.1 report with a heat number. A 2.2 only describes the product type, not your metal.
Three layers of proof: ingot, bar, finished piece
Proof can exist at three points between the melt and the polished piece. The ingot is the large block the mill melts first. It is then worked down into bar, and the bar is machined into jewellery.
1 / 3
Ingot
- The question it answers
- Is the chemistry inside F136?
- The document
- Ingot analysis, top and bottom
- In our published documents
- Attached to every mill report
2 / 3
Bar
- The question it answers
- Is the bar strong, sound and the right size?
- The document
- EN 10204 3.1 mill report
- In our published documents
- Four reports from 2014–2018, plus one for the 2026 heat
3 / 3
Finished piece
- The question it answers
- Is the jewellery still the same alloy?
- The document
- Independent lab report
- In our published documents
- Two reports, 2023 and 2025
The bar report is the minimum you should accept. The ingot and finished-piece layers add depth.
How to read a 3.1 certificate, point by point
Read every 3.1 report in the same order. The figure below shows where each point usually sits on the page. Each legend item tells you what to look for and what to do.
Anatomy of a 3.1 certificate
The 10 points to check
- Certificate issuer: check the producing mill’s name and letterhead. A reseller may add a cover page, but the mill’s own report must follow.
- Document type: find “EN 10204 3.1”, printed clearly. A 2.1 or 2.2 is not enough.
- Especificaciones: confirm a named implant standard with its edition year, such as ASTM F136-13. ISO 5832-3 or B348 Grade 23 may appear as well.
- Heat number: match it on every page. It must be present, readable and identical each time.
- Product form and size: check the bar or wire diameter fits your pieces, such as 1.6 mm or 4.00 mm. Codes such as h7 describe how exact the diameter is.
- Condition and melting route: look for “annealed”. The melting method is often listed too.
- Chemical analysis: compare each measured value with the limits of the named standard. Values often come from the top and bottom of the ingot.
- Mechanical results: check tensile strength, yield strength and elongation against the minimums (F136: 860 MPa, 795 MPa, 10%). Each value needs its unit.
- Metallurgy: look for an alpha-beta structure, and a statement such as “free of alpha case” or “no continuous alpha case”.
- Signature and date: confirm a signature or stamp of the inspection representative, and a date.
How to read the numbers. Compare each value with the limit of the named standard. Read the value as printed, and do not round it yourself. Certificates may print decimals with a point or a comma. Values near a limit are explained below, with our own numbers.
Alpha case is a hard, brittle, oxygen-rich surface layer. “Free of alpha case” is the statement you want.
Other lines you may see
Mill reports often carry extra lines. They are not required by F136, but they tell you how carefully the bar was checked.
- Eddy-current test: an electrical test that finds cracks and flaws at or just below the bar surface, without touching it.
- 100% visual inspection: every bar was looked at, not only a sample.
- Annealing temperature and time: for example, 788 °C for 1 hour.
- “Cannot be reproduced except in full”: the mill wants you to read the whole report, never a single page.
- “Free of mercury and radioactive contamination”: a standard statement about the material.
Tremun’s certificates, opened line by line
Our página de certificados publishes the documents below. One document shows a packing list with eight heats of bar. Full mill reports are published for the four 1.6 mm heats, from 2014 to 2018. The 2026 heat, 30553DG, has its own declaration with the full mill report attached. Two independent lab tests on finished pieces complete the set.
You can read the original reports in full, with their issuers, on our certificates page. The table gives the key values from each report, copied as printed. Where a report gives top and bottom values, we show a range.
Our published results
| Heat or test | Year | Form, diameter | O % | Fe % | Al % | V % | Tensile MPa | Yield MPa | Elongation % |
|---|---|---|---|---|---|---|---|---|---|
| Heat 8V37036 | 2014 | Bar, 1.6 mm | 0.11– | 0.16– | 6.04– | 3.90– | 1001 | 924 | 11 |
| Heat PVD3696 | 2016 | Bar, 1.6 mm | 0.089– | 0.09– | 6.27 | 3.91– | 1059 | 913 | 14 |
| Heat PVD4439 | 2018 | Bar, 1.6 mm | 0.05 | 0.12 | 6.34– | 3.94– | 1039 | 888 | 14 |
| Heat PVD5351 | 2018 | Bar, 1.6 mm | 0.08– | 0.10– | 6.24– | 3.91– | 1062 | 905 | 17 |
| Lab test, finished pieces | 2023 | Finished pieces | 0.12 | 0.077 | 6.1 | 3.86 | Not tested | Not tested | Not tested |
| Lab test, finished pieces | 2025 | Finished pieces | 0.098 | 0.10 | 5.50 | 3.55 | Not tested | Not tested | Not tested |
| Heat 30553DG | 2026 (declaration) | Bar, 4.00 mm | 0.12– | 0.13 | 6.02– | 3.81– | 1043 | 907 | 22 |
| F136 limit | ≤0.13 | ≤0.25 | 5.50– | 3.50– | ≥860 | ≥795 | ≥10 |
Chemistry in weight per cent. Mill reports give ingot chemistry from the top and bottom of the ingot. Lab tests cover chemistry only. Heat 30553DG: declaration dated 2026; the heat was melted in 2024.
What the table shows
Every value sits inside F136. That holds for all seven sets of results, across two bar sizes.
- Oxygen: 0.05% to 0.13%. The F136 maximum is 0.13%.
- Iron: 0.077% to 0.17%. The maximum is 0.25%.
- Aluminium and vanadium: 5.50% to 6.41%, and 3.55% to 4.06%. Both stay inside their ranges.
- Carbon, nitrogen and hydrogen: carbon 0.009% to 0.028%, nitrogen below 0.016%, hydrogen 0.0009% to 0.006%. All are far below their limits. The full values are in the reports.
- Tensile strength: 1001 to 1062 MPa. The minimum is 860 MPa.
- Yield strength: 888 to 924 MPa. The minimum is 795 MPa.
- Elongation: 11% to 22%. The minimum is 10%.
The values also vary from heat to heat, as real measurements do.
Values near a limit. A limit is a pass line. A value exactly on the line conforms, and a value beyond it does not. Two of our values sit on a line. Aluminium is 5.50% in the 2025 lab test, the lower end of 5.50–6.50%. Oxygen is 0.13% at the top of the 2026 ingot, and 0.12% at the bottom. Both conform.
Every lab result also carries some measurement uncertainty. So do not adjust a printed value up or down yourself. For a lab test, the lab’s own conformity statement is what counts. Both of our lab reports state that the chemistry conforms to ASTM F136.
What else the reports record
The four older mill reports cover 1.6 mm bar, and the 2026 report covers 4.00 mm bar. Diameters carry tolerance codes: h7 on the 2016–2018 bar and h8 on the 4.00 mm bar.
The four 1.6 mm mill reports are signed and dated by the mill. The 2026 declaration attaches its mill report in full. Each mill report includes the ingot chemistry from top and bottom. The reports for heats 8V37036, PVD3696, PVD4439, PVD5351 and 30553DG state that the bar is free of alpha case. Several also record eddy-current testing.
The editions follow the rule above. The 2014 report cites ASTM F136-13 and ISO 5832-3 (1996). The 2016–2018 reports add ASTM B348 Grade 23. The 2026 documents cite ASTM F136-13 (R2021), B348 Grade 23 and ISO 5832-3:2021. Heat 30553DG was melted in March 2024, before F136-26 became active. So a 2026 declaration that cites the 2013 edition is what the edition rule predicts. The 2016 report also names a French and a British implant standard, NF S94-080-1 and BS 7252-3.
The 2026 report also lists a surface roughness of Ra ≤0.8 µm. That value describes the bar as delivered, before any polishing. It says nothing about the finish of a piece. Your loupe check covers that.
What an independent test of finished jewellery adds, and what it cannot prove
A mill report describes the bar. A finished-piece test checks the jewellery after every step in the workshop.
We sent finished jewellery to an independent laboratory twice, in 2023 and in 2025. The samples were finished posts, barbells and rings, not bar. The lab measured metals with an ICP-OES spectrometer, a standard instrument for metal analysis. Other instruments measured carbon, oxygen, nitrogen and hydrogen, partly at a partner laboratory. Both reports state that the chemistry conforms to Ti-6Al-4V under ASTM F136.
What it adds
- Chemistry after manufacture. Cutting, machining, threading and polishing all happen after the mill report. The test looks at the result.
- A check from outside the supply chain. The laboratory is neither the mill nor the supplier.
- Random samples. The 2023 report states that different samples were selected at random.
- A second date. The two tests are two years apart, on different pieces.
What it cannot prove
- Polish and surface. A chemistry test does not look at the finish. Your loupe does.
- Nickel release. The EU nickel rule uses the EN 1811 test on the finished product. That is a different test.
- Strength. These lab tests measure chemistry only. Strength comes from the bar report.
- A specific piece. The lab states that its values refer to the samples tested. They do not show where one piece in your hand came from.
How to judge any finished-piece test. Check five things. The sample must be finished jewellery, not bar. The report must name the standard it checks against. It needs measured values, not only “conforms”. It needs a date and a signature. And it must be the complete report, with all its pages.
Traceability you can see
Documents only help if they connect to the metal. Here is how that link looks in our published documents.
The packing list. One box of Ti-6Al-4V ELI bar is listed by diameter and heat number. It holds eight heats in four diameters, from 1.6 mm to 4 mm.
The tagged bundles. Photos show the bar bundles with hand-written tags. Each tag gives the diameter, heat number, length and weight. For the four 1.6 mm heats, the same heat number then appears on the published mill report.
Copy into your supplier record
- A list by diameter. Each bar diameter linked to its heat numbers.
- A label on the metal. A tag or label with the heat number on each bundle or lot.
- A full 3.1 report for each heat you rely on. Complete and unaltered.
- A link to your order, if offered. Some suppliers can name the heat behind each order.
- Finished-piece results. Lab results on finished jewellery, if the supplier has them.
Then keep your own record. For each delivery, note the date, supplier, product and the documents you received. A simple spreadsheet is enough. If a client or an inspector raises a question, you can answer with a document.
Warning signs: when to refuse a certificate
Each sign below means the document cannot show what it should. None of them proves fraud on its own.
- No heat number, or different numbers on different pages.
- No named specification. Only “titanium”, “G23” or “implant grade”.
- The wrong specification. B348 bar presented as F136 material.
- The wrong document. A 2.1 or 2.2, or a simple “certificate of conformity”.
- No validation. No signature, stamp or date.
- An altered copy with no original. Re-typed lines, changing fonts, cropped or blurred pages.
- No measured values. Only “conforms” or “pass”, or numbers without units.
- Values copied from the standard. Every element matches the printed maximum or minimum.
- One old document for everything. The same undated report for every size and every year.
One warning sign is enough to pause the order. With two or more, send your questions before any further order.
Our certificates
Our titanium certificates are on our certificates page. Ingot chemistry, five mill reports and two lab tests on finished pieces are there. The page is open before you register.
Open our certificatesQuestions to send a supplier
Copy into an email
- Which standard does your titanium meet: ASTM F136, ISO 5832-3, or both?
- Can you send the complete, unaltered 3.1 report for this material?
- Can I read your certificates before I order?
- How do your certificates relate to the pieces in my order?
- Does every titanium product page state the standard?
- Are any parts made from another metal, such as links or coatings?
- What finish do you apply to tissue-contact surfaces?
- Are posts internally threaded or threadless, and are pieces autoclave safe?
- What warranty covers manufacturing defects?
- Do you have independent test results on finished pieces?
Question 6 matters for chains and coated pieces. Only the titanium part can be F136. Any other metal or coating needs its own documentation.
Expect written answers before you place an order.
What we publish. Here is how Tremun answers all ten questions.
- Question 1: our titanium is ASTM F-136 (Ti-6Al-4V ELI), sourced in France.
- Question 2: we publish five complete mill reports, with all their pages and the ingot chemistry.
- Question 3: yes. Our certificates page is open before you register.
- Question 4: our certificates cover our titanium stock, not individual orders. Today we cannot tell you which heat covers a specific order. Ask us and we will send any report on that page as a complete PDF.
- Question 5: every titanium product page names the standard: ASTM F-136.
- Question 6: product pages declare any other material. Chains with niobium links and pieces with a PVD coating state it.
- Question 7: product pages name the finish where one is stated, such as “Mirror polish”.
- Question 8: product pages state the thread type and “Autoclave safe”.
- Question 9: a lifetime warranty covers every piece.
- Question 10: yes, from 2023 and 2025.
The 60-second check before you order
Before you order · 60 seconds
Run this on any supplier page.
- Standard named on the page? An implant standard, such as ASTM F136, on the product page.
- Grade complete? “Ti-6Al-4V ELI”, not “titanium” or “surgical titanium” alone.
- Certificates visible? A certificates page, or a clear statement that they are supplied.
- Certificate type 3.1? Open one. Look for “EN 10204 3.1” and a heat number.
- Finish named? “Mirror polish” or another stated finish.
- Thread type? Internally threaded or threadless.
- Other metals? Links, coatings or settings declared separately.
You need “yes” for checks 1 to 4, plus at least one of checks 5 to 7. Then order a first batch and inspect it with a loupe. Otherwise, send your questions first.
What the certificate does not cover: surface and finish
En APP minimum standard for initial jewellery makes this point itself. A certificate shows the grade of the material, not the origin of each piece. APP also advises buying from a reputable supplier.
The certificate also stops at the bar. Even a lab test on finished pieces checks chemistry, not finish. APP expects a mirror polish on every surface that touches tissue. Look for no tool marks, sharp edges or leftover polishing paste. Posts should be internally threaded or threadless.
On arrival · 10× loupe
Loupe check
- Post ends: smooth and rounded. Reject a rough edge or a sharp end.
- Thread condition: clean and even. Reject metal flakes or dark paste.
- Tissue-contact surfaces: an even mirror finish. Reject tool marks or small holes.
- Stone settings: no polishing paste left around stones or prongs.
- Top fit: the top threads in smoothly and sits flush. Reject it if it sticks or moves.
Print this list and keep it with your stock.
Check a few pieces from each new delivery, not only the first order. Write the result in your supplier record, next to the delivery date.
Where F136 meets APP standards and EU rules
France. The Public Health Code (article R1311-4) requires anything that goes through the skin to be sterile or sterilised before use. Initial jewellery is included. A ministerial order of 3 December 2008 requires studios to inform clients about risks, including allergy to jewellery metals. These rules do not name an alloy.
APP. The Association of Professional Piercers is an international non-profit association based in the United States. For initial jewellery, it accepts titanium to ASTM F136, ASTM F1295, ISO 5832-3 or ASTM F67.
REACH nickel rule. REACH Annex XVII, entry 27 covers posts inserted into pierced parts of the body. They must release less than 0.2 µg/cm² of nickel per week. The reference test is EN 1811. The rule measures release from the finished product, not the alloy name.
General Product Safety Regulation. Regulation (EU) 2023/988 applies since 13 December 2024. Each product needs a type, batch or serial number, plus the manufacturer’s contact details. Imported products also need the importer’s details. Check that each delivery carries this information before you sell or fit the pieces.
Every EU country. REACH and the GPSR apply in all EU member states, including Spain and France.
Legal status. “Implant grade” describes the bar standard, not the legal status of the piece. F136 is a professional standard, not a legal requirement.
How to explain F136 to a client
Clients often want absolute promises: no nickel at all, and no reaction ever. Avoid both.
At the counter
- “This titanium is made to ASTM F136. That specification was written for surgical implant material.”
- “Nickel is not an alloying element in this titanium. EU law also limits how much nickel piercing posts may release.”
- “The material comes with a 3.1 test report, and the supplier publishes it.”
- “An independent lab also tested finished pieces from this supplier.” (Use only if your supplier has such a test.)
- “No metal can promise zero reaction for every person. If you have a known metal allergy, tell me first.”
Avoid absolute health promises, and any approval or medical label the piece does not hold. Promise only facts you can show.
Demand the documents
Demand a named specification and a signed 3.1 report for each heat. Prefer a supplier who shows both openly, before you register. Then check the finish yourself, with a loupe, on every delivery.
Tremun titanium
Titanium with its certificates
- ASTM F-136 titanium (Ti-6Al-4V ELI), sourced in France.
- Mill reports and finished-piece lab tests, published.
- Decorative tops and rings, including Mirror polish finishes. Autoclave safe. Lifetime warranty on every piece.
More than 3,500 studios across Europe order from Tremun.
See Titanium JewelryRegister as a professionalFor professional piercers and studios. No minimum order. DHL Express from Barcelona, free over €350.
Preguntas frecuentes
What does ASTM F136 mean?
ASTM F136 is a material specification from ASTM International. It covers Ti-6Al-4V ELI titanium alloy for surgical implants. It fixes chemistry, minimum strength and the annealed structure. For piercers, it is a reference standard for initial jewellery titanium. Check the certificate for the metal, and check the finish of each piece yourself.
Is G23 titanium the same as ASTM F136?
Not automatically. Grade 23 is the trade name for Ti-6Al-4V ELI. ASTM F136 is the implant specification for that alloy. ASTM B348 Grade 23 is an industrial bar specification in the same alloy family, with different limits. For piercing jewellery, the certificate should name an implant standard, not “Grade 23” alone.
Is Grade 5 titanium acceptable for initial piercings?
No. Grade 5 usually means Ti-6Al-4V to ASTM B348, an industrial bar specification. It has no implant requirements, and the APP minimum standard for initial jewellery does not list it. Its limits are also wider: oxygen up to 0.20% (F136: 0.13%). Choose F136, ISO 5832-3, F1295 or F67 titanium with an EN 10204 type 3.1 report.
Does ASTM F136 titanium contain nickel?
Nickel is not an alloying element in ASTM F136. The alloy is titanium with aluminium and vanadium. Small traces of nickel can exist in almost any metal. EU law therefore limits nickel release instead of requiring zero nickel. Jewellery inserted into pierced skin must release less than 0.2 µg/cm² of nickel per week (EN 1811 test).
What does ELI mean?
ELI is a short code for “extra low” levels of small atoms in the metal: oxygen, nitrogen, carbon and hydrogen. These atoms make titanium harder, but more likely to break when it bends. F136 keeps them low, with oxygen at 0.13% or less. The metal then bends further before it breaks.
Is implant grade titanium expensive?
Yes, more than industrial titanium, because several costs combine. The alloy has tighter chemistry limits. Each melt is tested, and an authorised inspector signs the results. Jewellery then needs careful machining, precise threads and a mirror polish. Documented material also lets you answer client, inspector and insurer questions with documents.
Is a certificate enough to prove a piece is F136?
It is necessary, but not enough alone. A 3.1 report proves that the bar with one heat number met ASTM F136. It does not show that a given piece came from that bar. You also need a reliable supplier, consistent records and a clear link between certificates and stock. Then check the finish yourself.
Does a certificate citing an older F136 edition still count?
Yes, in most cases. The active edition is F136-26, published in February 2026. A certificate cites the edition valid when the bar was made. Older editions such as F136-13 remain common on mill reports. Check the measured values against the limits. Note the edition in your supplier records for future reference.
Is a value exactly at the limit acceptable?
Yes. A limit is a pass line, so a value exactly on it conforms. Oxygen at 0.13% conforms to F136, while 0.14% does not. Read the value as printed and do not round it. Every lab result carries some measurement uncertainty, so for a lab test, rely on the lab’s own conformity statement.
What does a lab test on finished jewellery prove?
It shows that finished pieces still have the right chemistry after cutting, machining and polishing. A good report names the standard, describes the samples and is signed and dated. It does not test the polish, nickel release or strength. It also cannot prove where one specific piece came from.